TSI OmniTrak Solution Core Module

User Calibration Guide
- Safety & Handling Notes
- Use calibration gases in a well-ventilated area.
- Do not exceed recommended gas concentrations or pressures.
- Secure gas cylinders upright and use appropriate regulators.
- Always wear PPE when handling compressed gases.
- ක්රමාංකනය කළ යුත්තේ කවදාද?
- Factory calibration – ensures initial accuracy
- වාර්ෂිකව
- Field calibration – compensates for real-world influences and aging over time
- At least daily or before each study
- More frequently if:
- Readings appear inaccurate or drift beyond expected range.
- It’s exposed to high VOC or CO2 environments, high humidity, dust, or pollutants.
- Regulatory or certification programs require it.
- Factory calibration – ensures initial accuracy
- අවශ්ය උපකරණ
- OmniTrak™ Core module calibration shroud (7590-84)

- Zero air source and/or span gas (see Recommended Gases and Concentrations table)
- Gas regulator for ≤ 0.3 L/min
- FEP lined tubing and regulator
- OmniTrak™ Smart Station
- OmniTrak™ Core module calibration shroud (7590-84)
- සැකසුම් උපදෙස්
- Start with the Core module paired to the Smart Station
- Attach the calibration shroud to the Core Module next to the probe
- Connect tubing from gas source to the calibration shroud
- On the Smart Station, tap the Connection menu (network icon in the top right corner). Find the Core module to calibrate and tap the kebab (three vertical dots) menu next to the module. Select ‘Calibrate’ and follow the instructions.
- ක්රමාංකනය
For the VOC-LOW sensor, Zero Calibration and Span Calibration are available. If both are performed, Zero Calibration should be completed before Span Calibration.
VOC Zero Calibration
- On the Smart Station Core Module Calibration screen:
- Select ‘VOC LOW’
- Tap ‘Calibrate’
- Tap ‘Zero’
- Apply the zero gas bottle. Tap ‘Continue.’
- Flow zero air into the sensor at ≤ 0.3 L/min
- Wait at least 2 minutes for reading to stabilize, then press ‘Apply’
- Wait for the confirmation message that the zero calibration was successful.

VOC Span Calibration
- On the Smart Station Core Module Calibration screen:
- Select ‘VOC LOW’
- Tap ‘Calibrate’
- Tap ‘Span’
- Flow VOC span gas (i.e., 10,000 ppb isobutylene) into the sensor at ≤ 0.3 L/min
- Wait at least 2 minutes for reading to stabilize, then press ‘Apply’
- Enter Span Gas Concentration (ppb)
- Wait for confirmation message that the span calibration was successful.

CO2 Span Calibration
- On the Smart Station Core Module Calibration screen:
- Select ‘CO2’
- Tap ‘Calibrate’
- Tap ‘Span’
- Flow CO2 span gas (i.e., 400 ppb CO2) into the sensor at ≤ 0.3 L/min
- Wait at least 2 minutes for reading to stabilize, then press ‘Apply’
- Enter Span Gas Concentration (ppm)
- Wait for confirmation message that the span calibration was successful.
Post-Calibration Verification
- Flow gas again and confirm sensor is reading within +/- 10% of target
- Return to ambient air and confirm baseline VOC and CO2 return to expected values.
දෝෂගවේෂණය
| කලාපය | විය හැකි හේතුව | විසඳුම |
| ක්රමාංකනය අසාර්ථක විය | අස්ථායී කියවීම් | Let readings stabilize before retrying |
| VOC reading too high post-zero | Contamination in zero air source | Replace or verify zero air source |
| CO2 span fails | Incorrect gas concentration | Verify gas cylinder and regulator |
උපග්රන්ථය
Recommended Gases and Concentrations
| සංවේදක වර්ගය | ක්රමාංකන වර්ගය | Gas Used | අරමුණ |
|
VOC |
ශුන්ය | Air with zero VOCs/reference zero air, ≤ 0.3 L/min | Establish baseline free of ambient VOCs |
|
පරතරය |
The known VOC of interest, isobutylene in air is recommended. <10,000 ppb (10 ppm) recommended, ≤ 0.3 L/min | Ensure correct sensitivity to target VOCs | |
| CO2 | පරතරය | CO2 in air, 400 to 2000 ppm recommended,
≤ 0.3 L/min |
Ensure accurate readings at expected levels |
TSI Technical Support
Please contact TSI or your local channel partner to make service arrangements and to receive a Service Request number. To fill out an online Service Request form, visit TSI Incorporated’s webඅඩවියේ tsi.com/support.
TSI සංස්ථාගත
500 කාඩිගන් පාර
වෙරළview, MN 55126
Tel:……….+1 800-680-1220
……….+1 651-490-2860
Fax:………+1 651-490-3824
TSI සංස්ථාපිත - අපගේ පිවිසෙන්න webඅඩවිය www.tsi.com වැඩි විස්තර සඳහා.

TSI and the TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other countries’ trademark registrations.
- එක්සත් ජනපද දුරකථන: +1 800 680 1220
- එක්සත් රාජධානියේ දුරකථන: +44 149 4 459200
- ප්රංශය දුරකථන: +33 1 41 19 21 99
- ජර්මනිය දුරකථන: +49 241 523030
- ඉන්දියා දුරකථන: +91 80 67877200
- චීන දුරකථන: +86 10 8219 7688
- සිංගප්පූරු දුරකථන: +65 6595 6388
P/N 6018700 Rev. A US (10/7/2025)
©2025 TSI Incorporated Printed in U.S.A.
ලේඛන / සම්පත්
![]() |
TSI OmniTrak Solution Core Module [pdf] උපදෙස් 30771, 7590-84, OmniTrak, Module, OmniTrak Solution Core Module, Solution Core Module, Module |