1. හැඳින්වීම
The Aideepen LC-100 Car Coating Thickness Gauge is a precise instrument designed for non-destructive measurement of coating and paint thickness on metal surfaces. It automatically detects substrate properties using electromagnetic induction and eddy current principles, making it an essential tool for various applications including manufacturing, metal processing, chemical industry, commercial inspection, and laboratories.
This manual provides detailed instructions for the proper use, calibration, and maintenance of your LC-100 thickness gauge.
2 ප්රධාන ලක්ෂණ
- Simple and convenient operation with ultra-high precision resolution of 1μm.
- Supports FE (magnetic induction) and NFE (eddy current) coating measurements.
- Measurement range for automotive paint from 0-2000μm.
- Large screen display showing working mode, number of readings, and battery level.
- Precise integrated probe for automatic substrate property detection.
- Widely applicable in manufacturing, metal processing, chemical industry, commercial inspection, and laboratories.
සංදර්ශක සහ බොත්තම් කාර්යයන්:
- පරීක්ෂණ මාදිලිය: Automatic (AUTO), Magnetic induction (F), Eddy current (NFE)
- Measurement reading: Displays the current thickness measurement.
- Statistical display: Shows average, minimum, maximum, and total number of readings.
- Automatic shutdown instruction: Indicates when the device will power off automatically.
- ඒකක ස්විචය: Toggles between μm, mm, and mils.
- පසුබිම් ආලෝක ස්විචය: සංදර්ශක පසුබිම් ආලෝකය පාලනය කරයි.
- අඩු බල දර්ශකය: බැටරිය අඩු වූ විට ඇඟවීම්.
- Probe mode switch: Changes between AUTO, F, and NFE modes.
- Clear statistics: Resets statistical data.
- ශුන්ය ක්රමාංකනය: Initiates zero calibration.
- Switching average / minimum / maximum / number: Cycles through statistical views.
- විමර්ශනය: The sensor for measurement.
3. පිරිවිතර
| පරාමිතිය | වටිනාකම |
|---|---|
| නිෂ්පාදන නම | ආලේපන ඝණකම මැනීම |
| ආකෘතිය | LC-100 |
| ද්රව්ය | ABS |
| විභේදනය | 0.01mm / 1mil |
| මිනුම් පරාසය | 0-2000μm |
| මිනුම් නිරවද්යතාව | ±(3%+1μm) |
| බල සැපයුම | 2 x 1.5V AAA බැටරි (ඇතුළත් කර නැත) |
| මෙහෙයුම් උෂ්ණත්වය | -10℃ ~ +50℃ |
| පවතින වර්ණ | Black, Yellow, Red |
| නිෂ්පාදන මානයන් | Approx. 12.5cm (4.9in) length, 5.2cm (2in) width, 2.5cm (1in) thickness |
4. පැකේජ අන්තර්ගතය
පැකේජය විවෘත කිරීමෙන් පසු, කරුණාකර සියලුම අයිතම පවතින බවත් හොඳ තත්ත්වයේ පවතින බවත් සහතික කර ගන්න:
- 1 x LC-100 Coating Thickness Gauge
- 1 x උපදෙස් අත්පොත (මෙම ලේඛනය)
- 6 x Calibration Sheets (various thicknesses)
5. පිහිටුවීම
5.1. බැටරි ස්ථාපනය
- උපාංගයේ පිටුපස බැටරි මැදිරි කවරය සොයා ගන්න.
- කවරය විවෘත කිරීමට එය ලිස්සා හෝ ගලවන්න.
- නිවැරදි ධ්රැවීයතාව (+/-) සහතික කරමින් (2) 1.5V AAA බැටරි දෙකක් ඇතුළු කරන්න.
- බැටරි මැදිරි කවරය ආරක්ෂිතව වසා දමන්න.
Note: Batteries are not included in the package and must be purchased separately.
6. මෙහෙයුම
6.1. බලය සක්රිය / අක්රිය කරන්න
- To power on: Press and hold the 'AUTO OFF' button.
- To power off: The device will automatically shut down after a period of inactivity to conserve battery. You can also manually power off by pressing and holding the 'AUTO OFF' button.
6.2. ඒකක මාරු කිරීම
Press the 'um/mm/mils' button to cycle through the measurement units: micrometers (μm), millimeters (mm), and mils.
6.3. Measurement Modes (Probe Mode)
The LC-100 automatically detects the substrate type (magnetic or non-magnetic). The display will show 'Fe' for magnetic substrates (e.g., steel, iron) and 'NFE' for non-magnetic substrates (e.g., aluminum, copper).
- ස්වයංක්රීය මාදිලිය: The gauge automatically identifies the substrate material (Fe or NFE) and selects the appropriate measurement method. This is the default and recommended mode for general use.
- F Mode (Magnetic Induction): Used for measuring non-magnetic coatings on magnetic metal substrates (e.g., paint on steel).
- NFE Mode (Eddy Current): Used for measuring non-conductive coatings on non-magnetic metal substrates (e.g., paint on aluminum).
To manually switch between modes, press the 'AUTO F/NF' button.
6.4. මිනුම් ගැනීම
- Ensure the device is powered on and the desired unit is selected.
- Place the probe firmly and perpendicularly onto the coated surface you wish to measure.
- The measurement reading will appear on the display.
- Lift the probe from the surface to take another measurement.
6.5. Statistical Display
The gauge can display statistical data from multiple readings. Use the dedicated button (often combined with zero calibration or mode switch, refer to the display diagram) to cycle through Average, Minimum, Maximum, and Number of readings (NUM).
To clear statistical data, press the 'Clear Statistics' button (indicated on the diagram).
7 ක්රමාංකනය
Regular calibration ensures the accuracy of your thickness gauge. The LC-100 supports both zero calibration and multi-point calibration.
7.1. ශුන්ය ක්රමාංකනය
Zero calibration is performed on an uncoated metal substrate to establish a baseline.
- Ensure the gauge is in the correct mode (Fe or NFE) for the substrate you are calibrating.
- Place the probe firmly on the bare, uncoated metal substrate (e.g., the provided standard aluminum base or a known bare metal surface).
- Press the 'ZERO CAL' button. The display should show '0.00' or a value very close to zero.
- If the reading is not zero, repeat the process.
7.2. Multi-Point Calibration
Multi-point calibration uses calibration sheets of known thicknesses to ensure accuracy across the measurement range.
- Perform a zero calibration first (as described above).
- Select a calibration sheet (e.g., 0.05mm). Place it on the bare metal substrate.
- Place the probe firmly on the calibration sheet.
- Adjust the reading using the appropriate buttons (refer to the device's specific interface, usually 'up'/'down' or 'CAL' button with subsequent adjustments) until it matches the known thickness of the calibration sheet.
- Repeat this process for several other calibration sheets (e.g., 0.25mm, 0.50mm, 1.00mm, 2.00mm) to ensure accuracy across the range.
Always calibrate the device before critical measurements or if you suspect inaccuracies.
8. නඩත්තු කිරීම
- පිරිසිදු කිරීම: මෘදු වියළි රෙද්දකින් උපාංගය පිස දමන්න. උල්ෙල්ඛ පිරිසිදු කරන්නන් හෝ ද්රාවක භාවිතා නොකරන්න.
- පරීක්ෂණ සත්කාර: Keep the probe clean and free from dust, dirt, or debris. Avoid scratching the probe surface.
- ගබඩා: Store the gauge in a dry, cool place, away from direct sunlight and extreme temperatures. If storing for extended periods, remove the batteries to prevent leakage.
- බැටරි ආදේශනය: Replace batteries when the low power indicator appears on the display.
9. දෝශ නිරාකරණය
| ගැටලුව | විය හැකි හේතුව | විසඳුම |
|---|---|---|
| උපාංගය බලගන්වන්නේ නැත. | මිය ගිය හෝ වැරදි ලෙස ස්ථාපනය කර ඇති බැටරි. | Check battery polarity; replace with fresh AAA batteries. |
| වැරදි කියවීම්. | Needs calibration; dirty probe; incorrect measurement mode. | Perform zero and multi-point calibration. Clean the probe. Ensure correct Fe/NFE mode is selected or use AUTO mode. |
| සංදර්ශකය අඳුරු හෝ දිලිසෙන සුළුය. | අඩු බැටරි. | බැටරි ප්රතිස්ථාපනය කරන්න. |
| No reading or 'Err' message. | Probe not firmly on surface; surface not metal; measurement out of range. | Ensure firm contact. Verify the surface is metal. Check if the coating thickness is within 0-2000μm. |
10 පරිශීලක ඉඟි
- For consistent readings, always place the probe perpendicularly to the surface and apply gentle, even pressure.
- When measuring curved surfaces, take multiple readings at different points to ensure accuracy.
- Avoid taking measurements on dirty, rusty, or uneven surfaces, as this can affect accuracy.
- Keep the calibration sheets in good condition and store them properly to ensure they remain accurate for future calibrations.
- If measuring multiple vehicles or different types of metal, it's good practice to perform a quick zero calibration on a known bare metal surface before starting.
11. වගකීම් සහ සහාය
වගකීම් තොරතුරු, තාක්ෂණික සහාය හෝ සේවා විමසීම් සඳහා, කරුණාකර ඔබේ සිල්ලර වෙළෙන්දා හෝ නිෂ්පාදකයා සෘජුවම අමතන්න. මිලදී ගැනීම පිළිබඳ සාක්ෂියක් ලෙස ඔබේ මිලදී ගැනීමේ රිසිට්පත තබා ගන්න.
Please note that this device is not intended for legal metrology or commercial transactions where certified accuracy is required unless explicitly stated by local regulations and verified by certified calibration services.