1. හැඳින්වීම
The Extech DO600 Waterproof ExStik II Dissolved Oxygen Meter is designed for precise measurement of dissolved oxygen levels. It displays oxygen concentration as a percentage of saturation (0 to 200%) or as a concentration in parts per million (ppm) or milligrams per liter (mg/L) (0 to 20.00 ppm/mg/L). Key features include adjustable altitude compensation, adjustable salinity compensation, and an analog bargraph for trend indication. This manual provides instructions for the proper setup, operation, and maintenance of your DO600 meter.
2. පැකේජ අන්තර්ගතය
ඇසුරුම් ඉවත් කිරීමෙන් පසු සියලුම අයිතම හොඳ තත්ත්වයේ පවතින බව තහවුරු කර ගන්න:
- Extech DO600 Dissolved Oxygen Meter
- DO ඉලෙක්ට්රෝඩය
- Protective Sensor Cap
- Spare Membrane Cap
- Electrolyte Solution
- Four 3V CR-2032 Batteries (pre-installed or included separately)
- 48-inch (1.2m) Neck Strap

Figure 1: Extech DO600 meter with its accessories, including the electrode, caps, electrolyte, and neck strap.
3. මීටර සංරචක

රූපය 2: ඉදිරිපස view of the Extech DO600 meter.
- LCD සංදර්ශකය: Shows dissolved oxygen readings, temperature, and other indicators.
- මාදිලිය/රඳවා තබා ගැනීමේ බොත්තම: Toggles between measurement modes and holds the current reading.
- CAL/RECALL Button: Initiates calibration and recalls stored data.
- සක්රිය / අක්රිය බොත්තම: Powers the meter on or off.
- ඉලෙක්ට්රෝඩය: Sensor for dissolved oxygen measurement.
- Protective Sensor Cap: Covers and protects the electrode.
4. පිහිටුවීම
4.1. බැටරි ස්ථාපනය
The DO600 meter uses four 3V CR-2032 batteries. These are typically pre-installed. If replacement is necessary:
- Twist off the battery compartment cap located at the top of the meter.
- Remove the old batteries, noting their polarity.
- Insert new CR-2032 batteries with the correct polarity.
- Securely replace the battery compartment cap to maintain the waterproof seal.
4.2. Electrode Preparation
Before first use or after prolonged storage, the DO electrode requires preparation:
- Unscrew the protective sensor cap from the electrode.
- Fill the spare membrane cap with the provided electrolyte solution. Ensure no air bubbles are trapped.
- Carefully screw the membrane cap onto the electrode, ensuring the membrane is taut and free of wrinkles.
- Allow the electrode to stabilize for approximately 15-30 minutes in the electrolyte solution before calibration.
5. මෙහෙයුම් උපදෙස්
5.1. බලය සක්රිය / අක්රිය කරන්න
ඔබන්න සක්රිය / අක්රිය button to power the meter on or off.
5.2 ක්රමාංකනය
Regular calibration ensures accurate readings. Calibrate the meter in air-saturated water or a known standard solution.
- Ensure the electrode is prepared as described in Section 4.2.
- Immerse the electrode in air-saturated water (e.g., tap water stirred vigorously for several minutes) or a DO standard solution.
- ඔබා අල්ලාගෙන සිටින්න කැල්/නැවත අමතන්න "CAL" තිරය මත දිස්වන තුරු බොත්තම ඔබාගෙන සිටින්න.
- The meter will automatically recognize the calibration point and adjust. Wait for the reading to stabilize and for the calibration indicator to confirm completion.
5.3. මිනුම් ගැනීම
To measure dissolved oxygen:
- මීටරයේ බලය.
- Immerse the electrode into the sample solution. Ensure the membrane is fully submerged and free of air bubbles.
- Gently stir the electrode in the sample to ensure proper flow across the membrane.
- Wait for the reading on the LCD display to stabilize. The display will show both dissolved oxygen and temperature.
- ඔබන්න මාදිලිය/රඳවා තබා ගැනීම button to cycle between % saturation and ppm (mg/L) readings.
- To hold a reading, briefly press the මාදිලිය/රඳවා තබා ගැනීම button. Press again to release.

Figure 3: The Extech DO600 meter being used to measure dissolved oxygen in a laboratory.
5.4. Altitude Compensation
Dissolved oxygen saturation levels are affected by altitude. The DO600 allows for manual adjustment:
- Refer to the meter's full instruction manual for specific steps to enter and adjust altitude compensation (0 to 20,000 feet in 1,000 ft. increments).
5.5. Salinity Compensation
Salinity also influences dissolved oxygen levels. The DO600 provides adjustable salinity compensation:
- Consult the meter's comprehensive manual for detailed instructions on setting salinity compensation (0 to 50 ppt).
6. නඩත්තු කිරීම
6.1. Electrode Cleaning and Storage
To maintain electrode performance:
- එක් එක් භාවිතයෙන් පසු ආස්රැත ජලය සමග ඉලෙක්ට්රෝඩය සේදීම.
- If the membrane becomes fouled, gently clean it with a soft cloth or cotton swab. Replace the membrane cap if it is damaged or heavily contaminated.
- When storing, ensure the protective sensor cap is filled with electrolyte solution to keep the membrane hydrated.
6.2. බැටරි ප්රතිස්ථාපනය
Replace batteries when the low battery indicator appears on the display. Refer to Section 4.1 for battery installation instructions.
7. දෝශ නිරාකරණය
| ගැටලුව | විය හැකි හේතුව | විසඳුම |
|---|---|---|
| මීටරය ක්රියාත්මක නොවේ | මිය ගිය බැටරි; වැරදි බැටරි ස්ථාපනය. | බැටරි මාරු කරන්න; බැටරි ධ්රැවීයතාව පරීක්ෂා කරන්න |
| සාවද්ය හෝ අස්ථායී කියවීම් | Electrode not calibrated; Fouled membrane; Air bubbles on membrane; Insufficient electrolyte | Perform calibration; Clean/replace membrane; Remove air bubbles; Refill electrolyte |
| No reading displayed | Electrode not connected properly; Damaged electrode | Ensure electrode is securely attached; Replace electrode if damaged |
8. පිරිවිතර
| පරාමිතිය | වටිනාකම |
|---|---|
| ආකෘතිය | ඩොක්ස්නිය් |
| Dissolved Oxygen Range (Concentration) | 0 සිට 20.00 ppm (mg/L) |
| Dissolved Oxygen Range (Saturation) | 0 සිට 200.0% දක්වා |
| උෂ්ණත්ව පරාසය | Not explicitly stated for display, but upper rating is 50 ℃ |
| උන්නතාංශ වන්දි | 0 to 20,000 feet (in 1,000 ft. increments) |
| ලවණතා වන්දි | 0 සිට 50 දක්වා |
| බලශක්ති ප්රභවය | Four 3V CR-2032 batteries |
| මානයන් | 6.9 x 1.6 x 1.4 අඟල් (17.5 x 4.1 x 3.6 cm) |
| අයිතමයේ බර | අවුන්ස 3.87 (කිලෝ 0.11) |
| ජල ආරක්ෂිත ශ්රේණිගත කිරීම | Yes (implied by "Waterproof ExStik II") |

Figure 4: Physical dimensions of the Extech DO600 meter. Note: The image shows 7.7 inches, while product specifications state 6.9 inches. Refer to the latest product documentation for precise measurements.
9. නිෂ්පාදන වීඩියෝ
සපයා ඇති දත්තවල මෙම අයිතමය සඳහා විකුණුම්කරුගේ නිල නිෂ්පාදන වීඩියෝ කිසිවක් හමු නොවීය.
10. වගකීම් සහ සහාය
For warranty information, technical support, or to purchase accessories and replacement parts, please refer to Extech's official webවෙබ් අඩවියට පිවිසෙන්න හෝ ඔවුන්ගේ පාරිභෝගික සේවා දෙපාර්තමේන්තුව අමතන්න. ඔබට වැඩිදුර තොරතුරු සොයා ගත හැක Extech Store on Amazon.