1. හැඳින්වීම
This manual provides essential instructions for the proper use, maintenance, and troubleshooting of the Apera Instruments 201DJ-C AI3115 Double-Junction pH Electrode. Please read this manual thoroughly before using the electrode to ensure accurate measurements and extend its lifespan.
The Apera Instruments 201DJ-C AI3115 Double-Junction pH Electrode is specifically designed for long-term pH testing in challenging environments. Its robust construction makes it suitable for applications involving turbid water, wastewater, pools, spas, and general environmental monitoring. The double-junction structure is engineered to minimize clogging and contamination, which are common issues in such samples.
වැදගත් සලකා බැලීම්: This electrode is not recommended for measurements in solutions with very low ion concentrations (e.g., deionized or distilled water), highly alkaline solutions (pH > 12), strong corrosive liquids, or for prolonged measurements in high-temperature solutions (above 60°C).
The electrode is compatible with all pH meters that utilize BNC connectors. Please note that a temperature sensor is not integrated into this electrode. For accurate automatic temperature compensation (ATC), a separate temperature sensor is required, or the temperature must be manually entered into your pH meter.
2. නිෂ්පාදන අවසන්view
The Apera Instruments 201DJ-C AI3115 is a high-performance combined pH electrode featuring a double-junction design for enhanced stability and accuracy in complex samples.

Figure 1: Apera Instruments 201DJ-C AI3115 Double-Junction pH Electrode. This image shows the full electrode with its blue body and protective storage cap.
ප්රධාන ලක්ෂණ:
- මිනුම් පරාසය: 0-14 pH, 0-80 °C
- නිවාස ද්රව්ය: Polycarbonate for durability
- ප්රාචීරය: Double/Ceramic for reduced clogging
- යොමු ඉලෙක්ට්රෝඩය: Ag/AgCl for stable potential
- සම්බන්ධකය: BNC, ensuring wide compatibility

Figure 2: Key features of the 201DJ-C Double-Junction pH Electrode, highlighting its durable, accurate, stable, and fast response characteristics.
Electrode Design:
The double-junction design provides an additional barrier between the sample and the reference electrolyte, significantly reducing the risk of contamination and prolonging the electrode's life, especially in samples containing proteins, sulfides, or heavy metals.

රූපය 3: සමීප රූපය view of the pH sensor shield, designed to protect the sensitive glass bulb.
The electrode comes with a storage bottle containing a specialized solution to keep the sensor hydrated and in optimal condition when not in use.

Figure 4: The electrode is supplied with a storage bottle, ensuring the sensor remains in optimal status.
The BNC connector ensures broad compatibility with various pH meters.

Figure 5: The BNC connector, compatible with all BNC pH meters, provides a secure and standard connection.
3. පිහිටුවීම
- ඇසිරීම: Carefully remove the electrode from its packaging. Inspect for any visible damage.
- සකස් කිරීම: Remove the storage bottle from the electrode tip by twisting and pulling gently. Rinse the electrode tip with distilled or deionized water. Do not wipe the glass bulb, as this can create static electricity and affect readings.
- සම්බන්ධතාවය: Connect the BNC connector of the electrode to the corresponding BNC input on your pH meter. Ensure a secure connection.
- Temperature Sensor (if applicable): If your pH meter supports automatic temperature compensation (ATC) and you have a separate temperature sensor, connect it to your meter. Otherwise, you will need to measure the sample temperature manually and input it into the pH meter.
- කන්ඩිෂන් කිරීම: For optimal performance, immerse the electrode in a pH 7.00 buffer solution for at least 30 minutes before initial use or after prolonged storage. This helps to rehydrate the glass membrane.
- ක්රමාංකනය: Calibrate your pH meter with the electrode using at least two standard buffer solutions (e.g., pH 7.00 and pH 4.00 or pH 10.00). Refer to your pH meter's instruction manual for specific calibration procedures.
4. මෙහෙයුම් උපදෙස්
- මෙයට පිළියමක්: Before each measurement, rinse the electrode thoroughly with distilled or deionized water to remove any residual buffer or sample.
- ගිල්වන්න: Immerse the electrode tip into the sample solution, ensuring that the glass bulb and the junction are fully submerged. Stir the solution gently to ensure homogeneity and allow the reading to stabilize.
- කියවන්න: Once the reading on your pH meter stabilizes (usually indicated by a stable icon or a consistent reading), record the pH value.
- උෂ්ණත්ව වන්දි: If using ATC, ensure the temperature sensor is also immersed in the sample. If manual temperature compensation is used, measure the sample temperature and input it into the pH meter before taking the pH reading.
- භාවිතයෙන් පසු සේදීම: After each measurement, rinse the electrode thoroughly with distilled or deionized water.
- ගබඩා: Place the electrode back into its storage bottle containing the appropriate storage solution. Ensure the tip is fully immersed.
5. නඩත්තු කිරීම
5.1 Cleaning the Electrode
- නිතිපතා පිරිසිදු කිරීම: Rinse the electrode with distilled or deionized water after each use.
- General Contamination: For general dirt or deposits, soak the electrode in a warm (not hot) solution of mild detergent for 15-30 minutes, then rinse thoroughly with distilled water.
- Oil/Grease: Soak the electrode in ethanol or methanol for 15-30 minutes, then rinse thoroughly with distilled water.
- ප්රෝටීන් තැන්පතු: Soak the electrode in a 0.1% pepsin solution in 0.1M HCl for 15-30 minutes, then rinse thoroughly with distilled water.
- Inorganic Deposits: Soak the electrode in 0.1M HCl or 0.1M HNO3 for 15-30 minutes, then rinse thoroughly with distilled water.
After cleaning, recondition the electrode in pH 7.00 buffer for at least 30 minutes before recalibrating.
5.2 ගබඩා කිරීම
Always store the electrode with its tip immersed in the original storage solution (3M KCl solution or pH 4 buffer with KCl). Never store the electrode dry or in distilled/deionized water, as this will dehydrate the glass membrane and shorten its lifespan.
5.3 Electrode Lifespan
The lifespan of a pH electrode is typically 1-2 years, even with proper maintenance. Factors such as frequent use, harsh samples, and high temperatures can shorten its life. Replace the electrode when it can no longer be calibrated accurately or provides unstable readings after proper cleaning and conditioning.
6. දෝශ නිරාකරණය
| ගැටලුව | විය හැකි හේතුව | විසඳුම |
|---|---|---|
| Slow response or unstable readings | Dehydrated glass membrane; Clogged junction; Contaminated buffer; Air bubbles in electrode | Recondition electrode in pH 7 buffer; Clean electrode; Replace buffer; Gently tap electrode to remove bubbles |
| වැරදි කියවීම් | Improper calibration; Contaminated buffer; Electrode aging; Temperature not compensated | Recalibrate with fresh buffers; Replace buffers; Replace electrode; Ensure proper temperature compensation |
| No reading or 'Error' message | Loose connection; Broken electrode; Meter malfunction | Check BNC connection; Inspect electrode for damage; Test with another electrode or contact meter manufacturer |
| පාවෙන කියවීම් | Insufficient conditioning; Contaminated junction; Temperature fluctuations | Allow more time for conditioning; Clean electrode; Ensure stable sample උෂ්ණත්වය |
7. පිරිවිතර
| පරාමිතිය | වටිනාකම |
|---|---|
| ආදර්ශ අංකය | AI3115 |
| pH පරාසය | 0 - 14 pH අගය |
| උෂ්ණත්ව පරාසය | 0 - 80 °C |
| නිවාස ද්රව්ය | පොලිකාබනේට් |
| ප්රාචීරය | Double / Ceramic |
| යොමු ඉලෙක්ට්රෝඩය | Ag/AgCl |
| සම්බන්ධක වර්ගය | BNC |
| මානයන් (L x W x H) | 16 x 1.27 x 1.27 සෙ.මී |
| බර | ග්රෑම් 109 කි |
| නිෂ්පාදකයා | Shanghai Sanxin Instrumentation |
| මුල් රට | චීනය |
8. වගකීම් සහ සහාය
For warranty information and technical support, please refer to the documentation provided with your specific purchase or visit the official Apera Instruments website. General support and contact information can typically be found on the manufacturer's product pages.