1. හැඳින්වීම
This manual provides detailed instructions for the proper operation, maintenance, and troubleshooting of the INSTRUKART Lutron TU-2016 Turbidity Meter. Please read this manual thoroughly before using the device to ensure accurate measurements and safe operation.
The Lutron TU-2016 is designed for measuring turbidity in various water sources, including lakes, water tanks, rivers, and mangroves. It features a wide measurement range, high resolution, and a user-friendly interface, adhering to ISO 7027 standards.
2 ආරක්ෂිත තොරතුරු
- Always handle the meter and its components with care to prevent damage.
- උපාංගය අධික උෂ්ණත්වයන්ට, ආර්ද්රතාවයට හෝ සෘජු හිරු එළියට දිගු කාලයක් නිරාවරණය නොකරන්න.
- Ensure the battery compartment is properly sealed to prevent water ingress.
- Keep the exterior of the test bottle clean and dry before insertion into the meter.
- Use only specified standard solutions for calibration.
- Dispose of batteries and solutions according to local regulations.
3. පැකේජ අන්තර්ගතය
ඇසුරුම් ඉවත් කිරීමෙන් පසු, පහත ලැයිස්තුගත කර ඇති සියලුම අයිතම පවතින බවත් හොඳ තත්ත්වයේ පවතින බවත් තහවුරු කර ගන්න:
- 1 Unit of TU-2016 Turbidity Meter
- 0 NTU standard solution
- 10 NTU standard solution
- Empty testing bottle (sample cuvette)
- Hard Carrying case
- උපදෙස් අත්පොත
- කර්මාන්තශාලා ක්රමාංකන සහතිකය

රූපය 3.1: Contents of the TU-2016 Turbidity Meter package, neatly organized within its hard carrying case.

රූපය 3.2: The TU-2016 Turbidity Meter along with its essential accessories for measurement and maintenance.
4. නිෂ්පාදන අවසන්view
The Lutron TU-2016 Turbidity Meter is a portable, microprocessor-controlled device designed for precise turbidity measurements. Key features include:
- ISO 7027 Compliance: Designed to meet international standards for turbidity measurement.
- මිනුම් ඒකකය: NTU (Nephelometric Turbidity Unit).
- Wide Auto Measurement Range: 0 සිට 1,000 NTU දක්වා.
- අධි විභේදනය: 0.01 NTU (for 0.00 to 50.00 NTU range) and 1 NTU (for 50 to 1,000 NTU range).
- Unique Optics Structure: Enables accurate readings from low to high turbidity levels.
- පරිශීලක-හිතකාමී අතුරුමුහුණත: Four operation buttons and three calibration points for easy use.
- Jumbo LCD Display: Large display for clear and easy readability.
- Microprocessor Circuit: Ensures maximum accuracy and provides special functions.
- බැටරිය ක්රියාත්මක වේ: Ideal for field and on-site testing.
- දත්ත රඳවා ගැනීමේ කාර්යය: Freezes the displayed value.
- මතක මතකය: Records and recalls Maximum and Minimum readings.
- ස්වයංක්රීයව වසා දැමීම: බැටරි ආයු කාලය ඉතිරි කරයි.
- කල් පවතින නිර්මාණය: Heavy-duty and compact housing with a hard carrying case.

රූපය 4.1: ඉදිරිපස view of the TU-2016 Turbidity Meter, highlighting the display and control panel.

රූපය 4.2: The TU-2016 Turbidity Meter held in hand, illustrating its compact and portable design.
5. පිහිටුවීම
5.1 බැටරි ස්ථාපනය
The meter operates on six DC 1.5 V AAA (UM4) batteries. To install or replace batteries:
- මීටරයේ පිටුපස බැටරි මැදිරි කවරය සොයා ගන්න.
- Open the cover and insert six AAA batteries, observing the correct polarity (+/-).
- බැටරි මැදිරි කවරය ආරක්ෂිතව වසා දමන්න.
5.2 ආරම්භක බලය සක්රීය කිරීම
ඔබන්න බලය button to turn on the meter. The LCD display will illuminate and show a brief self-test before entering measurement mode.
6. මෙහෙයුම් උපදෙස්
6.1 Preparing a Sample
- Ensure the empty testing bottle (cuvette) is clean and dry. Wipe the exterior with a lint-free cloth.
- Fill the testing bottle with the water sample to be measured. Avoid air bubbles.
- Securely cap the bottle.
6.2 මිනුමක් ගැනීම
- එස් එක විවෘත කරන්නample compartment lid on the top of the meter.
- සකස් කළ s ඇතුල් කරන්නample bottle into the compartment, ensuring it is seated correctly.
- s වසන්නample මැදිරි පියන.
- ඔබන්න TEST/CAL button to initiate a measurement. The turbidity reading will appear on the LCD display.

රූපය 6.1: Opening the sample මැදිරි පියන.

රූපය 6.2: Inserting a sample bottle into the meter.

රූපය 6.3: The meter with the sample compartment open, ready for sample insertion.
6.3 දත්ත රඳවා ගැනීමේ කාර්යය
To freeze the current reading on the display, press the අල්ලා ගන්න බොත්තම. මුදා හැරීමට සහ සජීවී මිනුම් නැවත ආරම්භ කිරීමට එය නැවත ඔබන්න.
6.4 Memory Recall (MAX/MIN)
The meter can record and recall the maximum and minimum turbidity readings:
- ඔබන්න REC button once to display the maximum recorded value.
- ඔබන්න REC button again to display the minimum recorded value.
- ඔබන්න REC button a third time to exit memory recall mode.
6.5 ස්වයංක්රීයව වසා දැමීම
The meter features an auto shut-off function to conserve battery life. If no button is pressed for a certain period, the meter will automatically power off. This can be manually overridden by pressing the බලය උපාංගය අක්රිය කිරීමට බොත්තම.
7 ක්රමාංකනය
The TU-2016 features a three-point push-button calibration system. Regular calibration ensures the accuracy of your measurements. Use the provided 0 NTU and 10 NTU standard solutions for calibration.
- මීටරය සක්රිය කරන්න.
- Insert the 0 NTU standard solution into the sample මැදිරිය.
- ඔබා අල්ලාගෙන සිටින්න TEST/CAL button until the calibration mode is indicated on the display. Follow the on-screen prompts to calibrate the 0 NTU point.
- Remove the 0 NTU solution and insert the 10 NTU standard solution.
- Repeat the calibration process for the 10 NTU point.
- The third calibration point (if applicable) can be performed similarly with a higher NTU standard solution if available and required for your specific application.
- After successful calibration, the meter will return to measurement mode.
Note: Refer to the Factory Calibration Certificate for details on the initial calibration performed by the manufacturer.
8. නඩත්තු කිරීම
8.1 Cleaning the Meter
- මීටරයේ පිටත කොටස මෘදු, d රෙදි කඩකින් පිස දමන්න.amp රෙදි. උල්ෙල්ඛ පිරිසිදු කරන්නන් හෝ ද්රාවණ භාවිතා නොකරන්න.
- එස් තබා ගන්නample compartment clean and free of debris.
- Ensure the exterior of the test bottles are always clean and dry before use.
8.2 ගබඩා කිරීම
- When not in use, store the meter and its accessories in the provided hard carrying case.
- සෘජු හිරු එළියෙන් සහ අධික උෂ්ණත්වයන්ගෙන් ඈත්ව සිසිල් වියළි ස්ථානයක ගබඩා කරන්න.
- දිගු කාලයක් ගබඩා කරන්නේ නම්, කාන්දු වීම වැළැක්වීම සඳහා බැටරි ඉවත් කරන්න.
9. දෝශ නිරාකරණය
| ගැටලුව | විය හැකි හේතුව | විසඳුම |
|---|---|---|
| මීටරය ක්රියාත්මක නොවේ. | මිය ගිය හෝ වැරදි ලෙස ස්ථාපනය කර ඇති බැටරි. | නිවැරදි ධ්රැවීයතාව සහතික කරමින් බැටරි ප්රතිස්ථාපනය කරන්න. |
| වැරදි කියවීම්. |
|
|
| Display shows "Err" or similar error. | Measurement range exceeded or internal error. | සහතික කරන්න එස්ample is within 0-1000 NTU range. Power cycle the device. If error persists, contact support. |
10. පිරිවිතර
| පරාමිතිය | විස්තර |
|---|---|
| පරිපථය | Custom one-chip of microprocessor LSI circuit |
| ප්රදර්ශනය කරන්න | LCD ප්රමාණය: 41 mm x 34 mm |
| මිනුම් පරාසය | 0.00 to 50.00 NTU, 50 to 1,000 NTU (Auto range) |
| විභේදනය | 0.01 NTU (0.00 to 50.00 NTU), 1 NTU (50 to 1,000 NTU) |
| නිරවද්යතාව | ± 5 % F.S. or ± 0.5 NTU, whichever is greater |
| ආලෝක ප්රභවය | LED, 850 nm |
| අනාවරකය | ඡායාරූප ඩයෝඩය |
| සම්මත අනුකූලතාව | Meet ISO 7027 |
| ප්රතිචාර කාලය | තත්පර 10 ට අඩු |
| Sample Volume Needed | 10 මිලි |
| දත්ත රඳවා තබා ගැනීම | සංදර්ශක කියවීම කැටි කරන්න |
| මතකය නැවත කැඳවීම | උපරිම සහ අවම අගය |
| සංදර්ශකය එස්ampකාලය | දළ වශයෙන්. තත්පර 1 යි |
| බලය අක්රියයි | තල්ලු බොත්තම මගින් ස්වයංක්රීයව වසා දැමීම හෝ අතින් ක්රියා විරහිත කිරීම |
| මෙහෙයුම් උෂ්ණත්වය | 0 සිට 50 ° C |
| මෙහෙයුම් ආර්ද්රතාවය | RH 85% ට අඩු |
| බල සැපයුම | DC 1.5 V battery (UM4, AAA) x 6 PCs |
| Power Current (Standby) | ආසන්න වශයෙන්. DC 3.5 mA |
| Power Current (Testing) | ආසන්න වශයෙන්. DC 36 mA |
| බර | 320 g Approx. (0.71 lbs) |
| මානයන් | 155 x 76 x 62 mm (අඟල් 6.1 x 2.99 x 2.44) |
11. සහාය සහ වගකීම්
For technical assistance, troubleshooting beyond this manual, or warranty inquiries, please contact INSTRUKART customer support. Refer to your purchase documentation for specific warranty terms and contact information.
The product comes with a Factory Calibration Certificate, ensuring its initial accuracy and compliance with international standards.